Bulb Cannula Safety for Breast Fat Grafting

Sforza M., Martinez N., Araujo N., de Mezerville R., Castro JA.

Laboratory Study, published in Aesthet Surg J Open Forum (2020) — summary generated from the PubMed abstract.

Open my reading list
Level D· Scientific groundwork from lab and animal studiesEvidence level of this study

Evidence from laboratory and animal studies provides groundwork for understanding mechanisms and potential before human studies continue.

  • Level A · Stronger Clinical Evidence
  • Level B · Emerging clinical evidence with positive signals
  • Level C · Early human research exploring benefits
  • Level D · Scientific groundwork from lab and animal studies
  • Emerging · Emerging topic under active research
Read the A–D evidence level guide

This page is generated from the PubMed record. The Thai description is an automated summary of bibliographic fields and the abstract, not a full translation, and is not medical advice.

Study type
Laboratory Study
Journal
Aesthet Surg J Open Forum (2020)
Reported sample size
—
Source database
Europe PMC
PMID
33791641
PMCID
PMC7780445
DOI
10.1093/asjof/ojaa014
Citations
1

Abstract (original English)

Autologous fat transfer is a common technique to refine the contour of the breast after prosthetic augmentation or reconstruction, correcting remaining asymmetries by injecting previously harvested fat tissue with a cannula. Current procedures are often performed without visualization of the cannula at the delivery site and may require subsequent verification of the implant's integrity. The present paper aims to evaluate the safety of a new bulb tip cannula to be used during breast implant procedures for injecting fat adjacent to a breast implant that reduce the risk of damaging the implant. Two conventional cannulae and 3 bulb cannulae, which have an atraumatic distal tip, were tested in a simulated implant-puncture setting in 3 positions (at 0°, 45°, and 90° of incidence). A Tensile Tester (Instron, High Wycombe, UK) was used to apply force with each cannula device and record the amount of force applied in the attempt to penetrate the implants used, with shell layers having a variable thickness. No implant rupture was observed with the bulb tip cannulae regardless of size or the position in which the cannulae were pressed against the implants. The cannula opening was not impeded and tended to bend instead; 27% of the cases with the conventional lipo-cannulae caused an implant rupture. The bulb tip cannula could enhance the safety of the fat transfer procedure by ensuring no i

What this study does not prove

  • • This study does not prove SVF is an approved treatment or a replacement for standard care.
  • • This is preclinical work; animal or laboratory results cannot be applied to humans.